Página inicial > Produtos > High Pressure Homogenizer

NanoGenizer-Dual (Dual Pump)

Description: Portable Laboratory Dual-pump testing up to 25,000psi with electric power.
Código do item: NanoGenizer-Dual
    Selecione as informações desejadasX
  • Pressure:
    15,000psi 20,000psi 25,000psi 30,000psi
  • Flow Rate:
    120mL/min 180mL/min
  • Syringe Inlet&Outlet:
    10mL*2 20mL*2 50mL*2 100mL*2 200mL*2
  • Option: Inlet Stainless Cylinder:
    None 500mL 1000mL 2000mL 300mL-Glass-Jacketed
  • Option: Heat Exchanger:
    None HPE-100 HPE-150 HPE-150-3D
  • Voltage:
    220V
  • Warranty:
    Standard 1 year 2 year 3 year 4 year 5 year
  • IQ/OQ Installation and Operational Qualification:
    N/A IQ/OQ
Custo de envio:
para
Estimativa de Entrega:
Selecione o país do qual deseja enviar
+
-
Unidades
  • Descrição
  • How to Use
  • FAQ and Tips
  • How to Order
Description:
Portable Laboratory Dual-pump testing up to 30,000psi with electric power.
 
Application:
Nanodispersions, nanoemulsions and nanotube
Cell Disruption
Micro/nanoencapsulation in polymers, liposomes and oils
Solid dispersion and suspension, Deagglomeration
 
Performance:
Energy efficient, 1/2 energy loss of the most homogenizers
Small (Integrated design delivers the light weight and small dimensions)
Smart (Programming control systems confer diligent functions: Auto stop with time/volume)
Silent (Noiseless performance )
Strong (180 mL/min at 25,000 psi, 120 mL/min at 30,000 psi)
CE compliant and RoHs compliant
 
Specification
Catalog No. NanoGenizer-Dual
Flow Rate 180 mL/min at 25,000 psi, 120 mL/min at 30,000 psi
Min. Sample 30mL
Max. Pressure 30,000psi
Dimensions(cm) 72 × 43 × 33
Weight 60 kg (133 lbs)
Max. Temp. 80ºC (176ºF)
Power 220V
Cleaning Flush to clean
Application Nano emulsion, Fat emulsion, Liposome, Cell disruption, Nano dispersion, Deagglomeration
Features Digital display of pressure or pressure gauge
 
Standard Features
Parts Y-type Diamond interaction chamber with cooling option
Control System High pressure programming control systems®: Touch Screen, Speed control, Auto stop control by volume, time, pressure or temperature, Display of flow rate and time, Overload protection
Pressure Gauge Digital display on the touch screen
Inlet Type 1/4" HP coupling
Outlet Type 1/4" HP coupling
Inlet Reservoir 300mL S/S Cylinder
Outlet Reservoir S/S Pipe
Product Material 316L Stainless Steel, Tungsten Carbide, Viton, Teflon
Material Standard Pharmaceutical Grade, FDA, GMP
Warranty 1 year against any manufacturing defects
 
Option Features
Parts High Pressure Extruders, Homogenizing Valves Two Stages, Heat Exchanger
Control System Mechanical control: speed, start, stop control
Detector Pressure gauge, Pressure transducer up to 8 optional, Temperature transducer up to 8 optional
Outlet Type Tri-clamp or Luer
Inlet Reservoir 10mL, 20mL, 50mL, 100mL, 200mL Syringe, S/S Cylinder, Jacketed Glass Cylinder or Online
Outlet Reservoir 10mL, 20mL, 50mL, 100mL, 200mL, Syringe, S/S Cylinder, Jacketed Glass Cylinder or Online
Cylinder Titanium High Pressure Cylinder (Resistant to strong acid and base)
 
Varieties of Inlet and Outlet for NanoGenizer-Dual
Inlet:  Syringe; Open Container; Stainless Steel Cylinder
Outlet: Syringe; Open Container; Stainless Steel Cylinder
Inlet and outlet
 
Dissemble the seal


Assemble the seal

 
Troubleshooting:
Our design has minimized the occurrence of the air-blocking, chamber-blocking and overload.
However, the improper operation of the homogenizer can still induce the error and damage of the homogenizer.

Air Blocking
Symptom: There is no pressure or there is no sucking from inlet
Prevention: Do not run with empty inlet
Reason: There is air inside the pump
Solution:
1) In the Customer setup, push the button: Degas
2) Running 50mL water with 50% speed
3) Push the Ball in the inlet valve with a pin


Chamber Blocking
Symptom: Overload of the pressure
Prevention: Pre-treat the sample carefully; Reduce the speed
Reason: There is aggregates or particles in the sample; Too much speed
Solution:
1) Reverse flow the outlet valve with 100mL water with low speed
2) Reverse flow the Chamber with water from low to high speed

Overload
Solution:
1) Reduce the speed and close the power, wait 30 sec and restart again
2) Wash with water or your solvent
3) Running with clean and homo-disperse sample

The air blocking has rare damage for the Nanogenizer, however the chamber blocking is harmful to the pump because of overloaded pressure in an instant.
1) Pretreat the sample carefully, and always use clean and homo-disperse sample,
2) Adjust the pressure from low to high.
3) For safety, don not point the inlet or outlet of high pressure devices and equipment’s supplied by the Genizer, including high pressure homogenizer, extruder, interaction chamber and valve, to any personnel and object which may induce the safety issue and property loss.

Tips:
For facilitating the user applications, the two ports in the high pressure cylinder of NanoGenizer is exchangeable for inlet and outlet.


PRESSURE AND CHAMBER SELECTION FOR VARIOUS APPLICATIONS with NANOGENIZER
Cells Mammalian cell
Insect cell Bacteria cell Yeast cell Algae cell SPore
Sample CHO|Hela Sf9|Sf21|H5 E. Coli   Yeast Green algaea Ganoderma
Pressure 600-800bar   8500-11500psi 600-1000bar 8500-14500psi 800-1000bar 11500-14500psi 1200-1400bar 17500-20000psi 1200-1500bar 7500-21500psi 1300-1600bar 18500-23000psi
Chamber F20Y, F20Y-RT F20Y, F20Y-RT F20Y, F20Y-RT F20Y, F12Y F20Y, F12Y F20Y, F12Y












 
Emulsions Oil in water emulsions
Water in oil emulsions Oil emulsion Food Liposome (o/w) Liposome (w/o)
Sample Fat emulsion| Prostaglandin Cosmetics Dimethicone|Fuel Milk Doxorubicin DNA|lipopolyplex
Pressure ~20,000psi ~10,000psi ~20,000psi ~10,000psi ~20,000psi ~8,000psi
Chamber F20Y, F12Y F20Y F20Y, F12Y F20Y F20Y, F12Y F20Y
















 
Dispersions Pigments
Drugs/Creams Battery Particle aggregation Graphene Nanotubes
Sample Inks, Coatings Norfloxacin, Isoxyl Lithium/Solar cell Nano fibril Graphene Carbon Nanotubes
Pressure ~25,000psi ~20,000psi ~20,000psi ~25,000psi ~20,000psi ~20,000psi
Chamber F20Y, F12Y F20Y, F12Y F20Y, F12Y F20Y, F12Y F20Y, F12Y F20Y, F12Y
1. Order the unit from your Local Agent (Recommended)

2. Add the item into Cart and get the online quote from our website and confirm the order and payment accordingly (Recommended and Quick)
Get Quote and Order it Offline

3. Email us to get the quote

4. Order and pay online from the website if the Total Amount <$3000.00